VEGA VEGASON 71 - 1 … 75 - 1 User Manual
Page 38
Sensor optimization
38
71 - 1 … 75 - 1
Selection of sensor optimization
In mode 0 - 99 the sensor optimization 1 and 2 and
also the reset to the general parameter adjustment
can be carried out.
Programming example sensor optimization ..........
• select mode 0 - 99 in MODEFIELD
• program figure 1 in PARAMETER FIELD
• then store with STO-button
• operate MOD-button to enquire the sensor
optimization
Reset in mode 99, however with figure 0 in
PARAMETER FIELD.
Mode listing 1
Mode-no.
Mode description
Parameter
Page
(
= factory preset)
1 - 01
1 - 02
not used ....................................................
– – – –
1 - 03
Sensor type ...............................................
71 - 1 … 75 - 1
1 - 04
not used .....................................................
– – – –
1 - 05
-
indication, distance in m
1 - 06
-
indication, gain in dB
p. 41
Working range
1 - 07
-
start of working range ............................
0 m
1 - 08
-
end of working range ............................
6 /12 / 24 / 36 m
Product type
1 - 09
-
liquids / solids
......................
VEGASON 71 and 72 1 / 0
p. 42
......................
VEGASON 73 … 75
1 / 0
Liquid
1 - 10
-
vessel top flat / dished ..........................
1 / 0
p. 42
Solids
1 - 10
-
angle of repose .....................................
0 … 50 degrees
p. 43
1 - 11
-
particle size ...........................................
1 … 11 mm
Sensor optimization
0 - 99
0
- 01
1
1
- 01
27
1 - 99
0
MODE 0 - 99
Sensor optimization
Reset