G. Trouble Shooting
Problem
Probable Causes
Does not power up
•
Dead or defective battery
•
Broken wire from battery
snap to PCB
Won’t display DC
•
Open fuse
microamp readings
• Open test lead
•
Improperly connected to
circuit under test
All functions except
•
Very weak battery that will
ohms read high
not turn on the low battery
indicator on the LCD
AC Volts do not read
•
Very weak battery that will
not turn on the low battery
indicator on the LCD
Non-Contact Voltage
•
The non-contact voltage feature
does not work
works best when used to test a
single wire. Make sure the arrow
on the jaw is pointed at the wire.
The meter beeps when
•
This is a reminder to remove the
set on TEMP
test leads and insert the temper-
ature probe. If a probe is insert-
ed, this is an indication the probe
is open.
AC Amps does not read
•
Make sure the jaw is clamped
around a single wire and the
device connected to the wire is
turned on.
Application Notes
(Temperature Differential)
When measuring temperature there may be applica-
tions when knowing the differential between two mea-
surements is required. This can be accompished by
following these steps.
•
Remove the test leads from the input jacks. Insert
the K-type thermocouple probe into the input jack
on the 275. Set the 275 to the desired TEMP
function.
•
Touch the temperature probe sensor to the device
under test (T1).
•
When the reading stabilizes, press the REL but-
ton. The display will read “0” and T1 will be
stored as the reference value that the next mea-
surement will be made relative to.
•
Touch the temperature probe sensor to the device
under test (T2).
•
The 275 will display the differential temperature
T2-T1.