Veris Industries 275 Manual User Manual
Page 16

31
30
Application Notes (DC Volts)
The 275 will accurately measure rectified DC Voltages like
those encountered in furnaces and other appliances even
though many of these devices do not have output filtering
or other signal conditioning.
When measuring DC Voltage of a bat-
tery, the most accurate reading can be
attained by testing the battery under
load. To accomplish this, follow steps
1 through 4 shown on page 14 and
the following (with the battery in
holder and device turned on):
• Connect the red test lead from the
meter to the positive (+) terminal
of the battery.
• Connect the black test lead to the
negative (-) terminal of the battery.
F. Application Notes (AC Volts)
Disconnect power from the
terminal block, find the fuse
or circuit breaker that con-
trols the block and turn it
off.
Set up the meter following
the steps under
“Measurement Procedure”
on page 15. Then proceed with
the following:
•Connect the red test lead to
the hot side of the block and
the black lead to the neutral
side of the block. Reconnect
power to the block and read
the voltage on the meter. The
reading should be approximately 110V to
130V.
•
Disconnect power from the block and move
the red wire to ground. Reconnect power to
the block and read the voltage on the meter.
Typically less than 20V should exist from
neutral to ground. If 110V or above exists,
the block may be wired incorrectly.