beautypg.com

Units calib calib, Installation guide, Zero pressure calibration – Veris Industries PD SERIES Install User Manual

Page 3: Test, Pd series, Units calib test, Calib units

background image

ZL0055-0D

PAGE 3

©2012 Veris Industries USA 800.354.8556 or +1.503.598.4564 / [email protected]

07121

Alta Labs, Enercept, Enspector, Hawkeye, Trustat, Veris, and the Veris ‘V’ logo are trademarks or registered trademarks of Veris Industries, L.L.C. in the USA and/or other countries.

TM

PD SEriES

inStallation GUiDE

sPan Pressure calibration

current outPut calibration

The current output on the PD analog pressure gauge has been set at the factory.
These setting will not normally need adjustment. Perform these steps only if
necessary.

Units

Calib

Test

Current
measurement
device

Calib

Units

OR

Test

Calib

Units

Zero

VERIS INDUSTRIES

http://www.veris.com

1-800-354-8556

®

1-503-598-4564

Digital Pressure Gauge

zero Pressure calibration

Test

Calib

Units

Zero

VERIS INDUSTRIES

http://www.veris.com

1-800-354-8556

®

1-503-598-4564

Digital Pressure Gauge

Press ZERO then CALIB.
Hold both for 3 seconds.
Release when display reads zero
pressure.

The PD can be re-zeroed without affecting the span calibration. The gauge port must
be open with no pressure or vacuum applied.

Perform this step when pressure units show in the display (not mA). Zero calibration is retained
after unit is turned off.

Open gauge port.

test

To switch the PD into Test Mode, press and hold down the TEST button. The Test Mode
allows setup and testing of the output voltage without altering the system pressure.
To change the test output level, press and hold the TEST button, and press the CALIB
button while holding TEST. Each time the CALIB button is pressed, the test level
output increases by 2 mA.

Press and hold to enter Test Mode.

To change test output level, press
CALIB while holding TEST.

Units

Calib

Calib

1. Press to select pressure units.

2. Press and hold for 3 seconds, until display alternately

reads “CAL” and the calibration pressure (e.g. “CAL” and
“500” to calibrate at 500 psi).

3. Connect the gauge port to a pressure reference of known

accuracy. Apply pressure level selected in Step 2.

4. Press and hold the CALIB button for 3 seconds until

display reads “DONE.” Calibration is now complete.

Notes:
Only attempt Span calibration if the user has access to a pressure reference of known accuracy. Use

calibration equipment swith at least four times the gauge accuracy. Perform Zero calibration
before span calibration.

If the user tries to calibrate the unit beyond ±10% of the factory calibration, the display will read

“ERR” indicating an erroneous calibration.

To cancel calibration, press and release the CALIB button.
To restore factory calibration, repeat steps 1 and 2, then hold the ZERO button for 3 seconds, until

the display reads “FAC.”

1. Press to select the mA units.

2. Connect the leads to an accurate current measurement

device.

3. Press and hold for 3 seconds, until the display flashes

between “CAL” and “4.00.” Set the loop current to 4.00
mA

4. Press TEST to decrease the output by 0.01 mA. Press UNITS

to increase the output. While the output will change, the
display will not.

5. Press and hold for 3 seconds to accept the calibration at

4.00 mA. The display will change to “CAL” and “20.00.”

6. Repeat steps 3 through 5 to calibrate at 20 mA. Display will now read “DONE.”

Notes:
To cancel calibration changes, briefly press and release the CALIB button.
To restore factory calibration, repeat step 3 only, then hold the ZERO button for 3 seconds, until the

display reads “FAC.”