Universal Condenser (U-UCD8)

(Figs. 94 & 95)

<<

Attaching onto the UCD Adapter

>>

1. Using the Allen screwdriver, loosen the two setscrews @ on the IX-ADUCD

UCD adapter. (Fig. 94)

2. Push in the U-UCD8 universal condenser horizontally into the adapter

until the condenser’s positioning pin fits into the positioning groove on
the mount dovetail.

3. Tighten the two setscrews @ using the Allen screwdriver. (Fig. 94)

<<

Attaching onto the Illumination Column

>>

1. Rotate the condenser height adjustment knob ² to raise the condenser

to the upper limit position. (Fig. 95)

2. Loosen the condenser clamping knob ³. (Fig. 95)
3. Insert the mount of the UCD adapter into the mount dovetail of the

condenser holder (the condenser turret should be positioned on the
right hand side). Push the UCD adapter horizontally into the condenser
holder until the condenser’s positioning pin fits into the positioning groove
on the mount dovetail.

4. Tighten the condenser clamping knob ³. (Fig. 95)

<<

Optical Element Indices

>>

The optical element indices on the U-UCD-8 are not visible because it is
used upside down. Make your own reminder such as a list of the optical
elements inserted in the turret.
With the U-UCD8 (T2), however, the inserted optical elements can be
identified using the provided number stickers/indication plates.