beautypg.com

Other observation methods – Olympus IX71 User Manual

Page 36

background image

32

}A phase contrast objective, phase contrast optical element and the U-CT30 centering telescope (which is not required

when the U-BI90CT binocular observation tube is used) are required for phase contrast observation.

}If a DIC slider, analyzer and/or polarizer are engaged in the light path, disengage them.

OTHER OBSERVATION METHODS

5-1 Phase Contrast Observation (Using the IX2-ILL100 Column)

With the IX2-LWUCD or IX2-LWUCDA2 Condenser

Optical Element

Indication

Applicable Objectives

IX-PHL (small)

PhL

UPlanFl4XPh

IX-PHC (small)

PhC

CPlan10XPh, LCAch20XPh, CPlanFl10XPh

IX-PH1 (small)

Ph1

UPlanFl10XPh, UPlanFl20XPh, LCPlanFl20XPh, UPlanApo10XPh

IX-PH2 (small)

Ph2

UPlanFl40XPh, LCPlanFl40XPh, LCPlanFl60XPh, UPlanApo20XPh,
LCAch40XPh, LUCPlanFl40XPh, SLCPlanFl40XPh

IX-PH3 (large)

Ph3

UPlanFl100XO3Ph, UPlanApo40XOI3Ph, UPlanApo100XOI3Ph, PlanApo60XO3Ph

}Insert the optical element (small) in the 30 mm position and the optical element (large) in the 38 mm position.

For well observation, it is recommended to use the IX-PHC to obtain the phase contrast effect in a wide range of field of
view.

1

Phase Contrast Optical Elements and Applicable Objectives

Optical Element

Indication

Applicable Objectives

IX-PHL (small)

PhL

UPlanFLN4XPh

IX-PHC (small)

PhC

CPlanN10XPh, LCAchN20XPh, CPlanFLN10XPh

IX-PH1 (small)

Ph1

UPlanFLN10X2Ph, UPlanFLN20XPh, LUCPlanFLN20XPh

IX-PH2 (small)

Ph2

UPlanFLN40XPh, LUCPlanFLN40XPh, LUCPlanFLN60XPh,
LCAchN40XPh, UPlanFLN60XOIPh

IX-PH3 (large)

Ph3

UPlanFLN100XO2Ph

UIS2 Series

UIS Series

This manual is related to the following products: