beautypg.com

Tektronix CSA8000B User Manual

Page 85

background image

Acquiring Waveforms

CSA8000B & TDS8000B User Manual

3- 11

With an input signal connected, use the procedure that follows to autoset based
on the characteristics of the input signal. Autoset operates on the selected
channel only.

Overview

To autoset

Control elements and resources

Prerequisites 1.

The instrument must be installed with sampling modules
in place. Signals must be connected to channels. A
triggering source must be provided.

2.

At least one channel must be turned on (its front-panel
button lighted).

See the sampling-module user manuals for help with

installing sampling modules. See page 3--48 in this

manual for trigger setup information.

Execute 3.

Push the Autoset button to to execute an autoset on the
selected waveform.

If you use Autoset when one or more channels are
displayed, the instrument uses the selected channel for
horizontal scaling. Vertically, all channels in use are
individually scaled.

Note. Autoset can execute on live waveforms (either
channel or math) in the Main time base.

To Autoset the Instrument

This manual is related to the following products: