Perform the compensation – Tektronix CSA8000B User Manual
Page 38

Incoming Inspection
1- 20
CSA8000B & TDS8000B User Manual
Perform the Compensation
This procedure uses internal routines to verify that the instrument compensates
properly.
Equipment required
For sampling modules:
H 50 Ω terminations on all electrical module channels (Tektronix
part number 015-1022-xx).
H Dust covers on all optical module channels.
The sampling modules ship from Tektronix with the proper termina-
tions and dust covers installed.
Prerequisites
First, all sampling modules to be compensated must be installed as
outlined in their user manuals.
Second, power on the instrument and allow a 20 minute warm-up
before doing this procedure.
1.
Run the compensation routines:
a.
From the application menu bar, select Utilities, and then select Com-
pensation.
In the Compensation dialog box, the main instrument (mainframe) and
sampling modules are listed. The temperature change from the last
compensation is also listed. See Figure 1--6.
Click to select compensate
Click to start compensation
Choose all as targets
Figure 1- 6: Compensation dialog box