Triggering

CSA8000B & TDS8000B User Manual

3- 51

Overview

Control elements and resources

To use gated trigger (Cont.)

Enable gated

triggering

4.

In the Enhanced Triggering options section of the dialog
box, check Gated Trigger.

Check to

enable

Complete set up 5.

Attach an appropriate TTL-gating signal to the
TRIGGER GATE (TTL) rear-panel connector. Operation
is as follows:

H Triggering system will be disabled when the gating

signal is a TTL low, and instrument will not acquire.

H The triggering system will be enabled when the

gating signal is a TTL high, and the instrument will
acquire.

H There is an internal pull--up on the Gated Trigger

input such that if no drive signal is supplied or if the
input is left unconnected, triggering will be enabled
even if the Gated Trigger is selected in the Trig
Setup dialog box.

TRIGGER

GATE (TTL)

For more

information

6.

Press the Help button in the Trig Setup dialog box to
access the online assistance specific to triggering
commands. You can also read about key trigger
features in Keys to Using on page 3--40.

See page 3--167 for information on online help.

End of Procedure