Perform the diagnostics – Tektronix CSA8000B User Manual
Page 36

Incoming Inspection
1- 18
CSA8000B & TDS8000B User Manual
H One SMA 10X attenuator, such as Tektronix part number 015-1003-00.
H One or more (quantity to match number of electrical channels to compen-
sate) 50
Ω terminators, such as Tektronix part number 015-1022-01
H One 50 Ω terminator cap, such as Tektronix part number 011-0049-02
H One 80E00-series electrical sampling modules installed as outlined in its
User manual.
H One 80C00-series optical sampling module installed as outlined in its User
manual (optional; test only if purchased with/for your instrument).
H Mouse
H Keyboard
Perform the Diagnostics
The instrument Diagnostics use internal routines to confirm basic functionality
and proper adjustment.
Equipment required
None
Prerequisites
First, all sampling modules to be diagnosed must be installed as
outlined in their user manuals.
Second, power on the instrument and allow a 20 minute warm-up
before doing this procedure.
1.
Set up the instrument: From the application menu bar, select Utilities, and
then select Diagnostics. The Diagnostics dialog box displays. See below.