beautypg.com

Tektronix CSA8000B User Manual

Page 311

background image

Appendix B: Automatic Measurements Reference

CSA8000B & TDS8000B User Manual

B- 49

Table B- 7: NRZ Measurements - Amplitude (cont.)

Name

Definition

NRZ RMS Noise

One standard deviation of the amplitude variance sampled within a fixed width vertical slice
located at the center of the Eye Aperture at the High (logical 1) or Low (logical 0) levels.

RMS noise = High

σ or RMS noise = Lowσ

The Eye Aperture is adjustable and defaults to 20% of the NRZ bit time. The High or Low
selection for Noise At control in the Measurement Setup dialog instructs the measurement to be
performed on the logical 1 or 0 levels. See RZ Eye Aperture Parameters on B--62.

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2). See To Localize a Measurement on page 3--83.

This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
See Use a Waveform Database on page B--70.

For best results with this measurement:

H Perform a Dark Level compensation before taking this measurement if the source of the

measured waveform is an optical channel. See To Perform Dark-Level and User
Wavelength Gain Compensations on page 3--98.

H Optimize the vertical resolution before taking this measurement. See How to Optimize the

Vertical Resolution on page B--70.

NRZ Signal-to-Noise
Ratio

NRZ Signal-to-Noise is the ratio of the NRZ eye amplitude to the noise on either the High
(logical 1) or Low (logical 0) level. The data within the Eye Aperture is sampled, and the mean
of the histogram yields the High and Low levels. The noise is defined as one standard deviation
of the distribution within a fixed width vertical slice located at the center of the Eye Aperture.

Where High and Low are the logical 1 and 0 levels. See RZ Eye Aperture Parameters
on B--62.

The Eye Aperture is adjustable and defaults to 20% of the NRZ bit time. The High or Low
selection for Noise At control in the Measurement Setup dialog specifies that the measurement
be performed on the logical 1 or 0 levels. See To Localize a Measurement on page 3--83.

If enabled, measurement gates constrain the measurement region to the area between the Start
Gate (G1) and Stop Gate (G2).

This measurement requires the use of a waveform database. When this measurement is turned
on, it will automatically set the measurement system to use a waveform database if available.
See Use a Waveform Database on page B--70.

For best results with this measurement, perform a Dark Level compensation before taking this
measurement if the source of the measured waveform is an optical channel. See To Perform
Dark-Level and User Wavelength Gain Compensations on page 3--98.

S

ፒN Ratio =

(High

− Low)

High

σ

or

S

ፒN Ratio =

(High

− Low)

Low

σ

This manual is related to the following products: