OMWT-SOFT-OPC Server

Corrections N 55

Corrections

Why Use Corrections

Some devices, like a thermistor, need their data corrected in order to be meaningful.

A thermistor uses resistance to measure temperature. The relation between the

resistance and the temperature is non linear. A thermistor uses a set of correction

parameters to convert resistance to temperature. Other devices may need to correct

data values in a non linear way. OMWT-SOFT-OPC provides a way for the user to

correct the data from a sensor using a set of correction parameters or a lookup table.

Correction IDs

Tell OMWT-SOFT-OPC that you want a device to use a correction by specifying the

Correction ID in the OMWT-SOFT-OPC.INI file. Within the device section add a

line saying "CorrID=" and the number of the correction that you want OMWT-

SOFT-OPC to use. The Corrections are found in the Correction.INI file that you

write.

Pre-defined Thermistor

The thermistor made by the vendor Thermometrics has the part number RL0503-

5820-97-MS . This thermistor's parameters are predefined in OMWT-SOFT-OPC

and given the Correction ID 128. (In the device section of the OMWT-SOFT-

OPC.INI file, write "CorrID=128" to use this type.)

The parameters for the predefined thermistor are as follows:

NTC_DIVIDERES = 10000
NTC_NORMALRES = 10000
Fitting constants:
A = 3.3539264E-3
B = 2.5609446E-4
C = 1.9621987E-6
D = 4.6045930E-8