beautypg.com

Analog output digital triggering, Connecting analog output signals, Figure 7-2. analog output connections – National Instruments NI USB-621x User Manual

Page 71

background image

Chapter 7

Analog Output

NI USB-621x User Manual

7-4

ni.com

the buffer at a fast enough rate to keep up with the generation, the buffer
will underflow and cause an error.

Analog Output Digital Triggering

Analog output supports two different triggering actions:

Start trigger

Pause trigger

A digital trigger can initiate these actions on the USB-621x devices. Refer
to the

AO Start Trigger Signal

and

AO Pause Trigger Signal

sections for

more information about these triggering actions.

Connecting Analog Output Signals

AO <0..1> are the voltage output signals for AO channels 0 and 1.
AO GND is the ground reference for AO <0..1>.

Figure 7-2 shows how to make AO connections to the device.

Figure 7-2. Analog Output Connections

Load

Load

V OUT

V OUT

+

+

AO GND

AO 1

Analog Output Channels

M Series Device

AO 0

Channel 1

Channel 0

Digital

Isolators

Isolation

Barrier

(USB-6215

and USB-6218

devices only)