Table 5-1. analog input configuration, Ai ground-reference setting – National Instruments NI USB-621x User Manual
Page 52

Chapter 5
Connecting AI Signals on the USB-6210/6211 Devices
5-2
ni.com
Table 5-1. Analog Input Configuration
AI
Ground-Reference
Setting
*
Floating Signal Sources (Not
Connected to Building Ground)
Ground-Referenced Signal
Sources
**
Examples:
• Ungrounded thermocouples
• Signal conditioning with
isolated outputs
• Battery devices
Example:
• Plug-in instruments with
non-isolated outputs
Differential
Single-Ended—
Non-Referenced
(NRSE)
Single-Ended—
Referenced (RSE)
*
Refer to the
Analog Input Ground-Reference Settings
section of Chapter 4,
, for descriptions of the RSE,
NRSE, and DIFF modes and software considerations.
**
Refer to the
Connecting Ground-Referenced Signal Sources
section for more information.
+
–
+
–
AI+
AI–
AI GND
Signal Source
DAQ Device
+
–
+
–
AI+
AI–
AI GND
Signal Source
DAQ Device
Vcm
+
–
+
–
AI
AI SENSE
AI GND
Signal Source
DAQ Device
+
–
+
–
AI
AI SENSE
AI GND
Signal Source
DAQ Device
Vcm
+
–
+
–
AI
AI GND
Signal Source
DAQ Device
Ground-loop potential (V
A
– V
B
) are added
to measured signal.
NOT RECOMMENDED
for the
USB-6210/6211
+
–
+
–
AI
AI GND
V
B
V
A
Signal Source
DAQ Device