rhythm analysis

...................... See “

SMART

analysis.”

send (data)

............................. A feature of the HeartStart FR2+ that allows use of its infrared (IR)

communications port to send data directly to another FR2+ or a computer
running HeartStart Event Review Web software.

sensitivity

................................ A measure of the ability of the HeartStart FR2+ to reliably detect and identify

shockable heart rhythms.

setup

........................................ The settings of all programmable operating parameters of the HeartStart

FR2+. The factory default setup can be modified using the M3864A training
& administration pack.

shock series

........................... One or more shocks, each separated by no more than a preset time

(programmed Protocol Timeout). After completion of a shock series, the
HeartStart FR2+ automatically pauses for CPR.

shockable rhythm

.................. Ventricular fibrillation and certain ventricular tachycardias associated with

sudden cardiac arrest.

shock waveform

..................... See “

SMART

biphasic waveform.”

SMART analysis

...................... The proprietary algorithm used by the FR2+ to analyze the patient’s heart

rhythm and determine whether a shock is advised.

SMART biphasic
waveform

............................

The patented, low-energy defibrillation shock waveform used by the FR2+. It

is an impedance-compensated biphasic waveform with 150 Joules, nominal,
delivered to a 50 ohm load. When delivered via FR2 infant/child
reduced-energy defibrillator pads, the energy is attenuated to 50 Joules,
nominal.

specificity

................................ A measure of the ability of the HeartStart FR2+ to reliably detect and identify

non-shockable heart rhythms.

standard battery

...................... The M3863A battery, 12 VDC, 4.2 Ah, lithium manganese dioxide,

disposable, long-life primary cell.

standby mode

........................ The operating mode of the HeartStart FR2+ when a battery has been

installed, and the unit is turned off and ready for use when needed. Shown by
flashing black hourglass on the Status Indicator.

status indicator

...................... This is a special window in the upper right-hand corner of the front panel of

the HeartStart FR2+ that lets you know the status of the device.

sudden cardiac
arrest

...................................

The sudden cessation of the heart’s pumping rhythm.