Philips M3860A User Manual
Page 55
S e t u p a n d A d v a n c e d M o d e F e a t u r e s
6-7
P
h
ili
ps
Me
dic
a
l
Sy
s
tem
s
6
Use the infrared communications feature of the FR2+ to receive the
revised setup from another FR2+. (Instructions are provided later in this
chapter.)
Use the infrared communications feature of the FR2+ to receive the
revised setup from a computer running HeartStart Event Review
software. (Instructions are provided with the HeartStart Event Review
software.)
C A U T I O N
:
Alteration of the factory default setup of the FR2+ can affect its
performance and should be performed under the authorization of your
Medical Director. Modifications to device operation resulting from changes to
the default settings should be specifically covered in user training.
See the tables describing the various setup parameters at the beginning of
this chapter and Appendix D, “Glossary of Terms,” for definitions of setup
items.
Receiving setup
This method uses the infrared communications feature of the HeartStart
FR2+ to receive setup directly from one HeartStart FR2+ to another (which
must have the training & administration pack installed in it) or from a
computer running HeartStart Event Review software. (See instructions
provided with HeartStart Event Review.) To receive setup from another
FR2+, follow these steps:
1.
Locate the infrared communications port on each HeartStart FR2+ and
line them up with one another, so that the infrared “eye” in each one has
an uninterrupted view of the “eye” in the other. (See the diagram on the
inside front cover.) The two devices should be no more than 1 meter
apart.
2.
Make sure the “sending” FR2+ has the training & administration pack
installed and is ready to send. (See the M3864A training &
administration pack Reference Guide for instructions.)
3.
Select
RECEIVE
SETUP
from the setup menu:
4.
A new screen comes up. Until the two HeartStart FR2+ devices are
properly positioned, the screen displays
READY
TO
RECEIVE
and
prompts you to check the sending FR2+.
5.
Setup data are automatically transferred as soon as the infrared ports
are correctly aligned.
6.
If you select
EXIT
before the transfer is complete, the revised setup will
not be received. When the transfer is complete, the screen on the