beautypg.com

Philips M3860A User Manual

Page 75

background image

Te c h n i c a l S p e c i f i c a t i o n s

B-7

P

h

ili

ps

Me

dic

a

l

Sy

s

tem

s

B

(Optional) M3848A FR2+ rechargeable battery

(Optional) M3849A charger

M3870A and DP2/DP6 defibrillator pads

category

nominal specifications

Battery Type

11.3 VDC, 6.5 Ah, lithium ion. Rechargeable cell using the M3849A charger.

Capacity

When freshly charged and used at 77° F (25° C) , provides a minimum of 80
(typically 100) shocks, or 3.5 hours (typically 5 hours) of ECG display time only,
before recharging is indicated.

Status Indicators

Good battery: bar graph on display screen indicating remaining power level.
Low battery:

flashing red X on the front panel of the FR2+ (When low battery

indicator appears, there is still enough energy to deliver 9 shocks plus 15 minutes
of ECG display time).
Dead battery: solid red X on the front panel of the FR2+.

Storage Temperature

32° to 109° F (0° to 43° C).

Standby Life
(after installation)

6 months when installed fully charged in a defibrillator labeled FR2+.

category

nominal specifications

Application

For use with M3848A FR2+ rechargeable battery only.

Power Requirements

100 to 240 VAC, 47 to 63 Hz, 30 Watts

Environmental
Requirements

32° to 122° F (0° to 50° C).
0% to 95% relative humidity (non-condensing).

Conformance Testing

International: EN60335-1:1994 Class 1
North America: UL 1310 Class 2

category

nominal specifications

Pads, Cable, and Connector

Disposable and self-adhesive. DP2/DP6 adult defibrillator pads have a nominal
active surface area of 100 cm

2

each and are provided in a sealed package with an

integrated 122 cm (48 inch), typical, cable and connector. M3870A FR2
infant/child reduced-energy defibrillator pads have a nominal active surface area of
44 cm

2

each and are provided in a sealed package with an integrated 122 cm (48

inch), typical, cable and connector incorporating attenuating electronics.

This manual is related to the following products: