Te c h n i c a l S p e c i f i c a t i o n s
B-7
P
h
ili
ps
Me
dic
a
l
Sy
s
tem
s
B
(Optional) M3848A FR2+ rechargeable battery
(Optional) M3849A charger
M3870A and DP2/DP6 defibrillator pads
category
nominal specifications
Battery Type
11.3 VDC, 6.5 Ah, lithium ion. Rechargeable cell using the M3849A charger.
Capacity
When freshly charged and used at 77° F (25° C) , provides a minimum of 80
(typically 100) shocks, or 3.5 hours (typically 5 hours) of ECG display time only,
before recharging is indicated.
Status Indicators
Good battery: bar graph on display screen indicating remaining power level.
Low battery:
flashing red X on the front panel of the FR2+ (When low battery
indicator appears, there is still enough energy to deliver 9 shocks plus 15 minutes
of ECG display time).
Dead battery: solid red X on the front panel of the FR2+.
Storage Temperature
32° to 109° F (0° to 43° C).
Standby Life
(after installation)
6 months when installed fully charged in a defibrillator labeled FR2+.
category
nominal specifications
Application
For use with M3848A FR2+ rechargeable battery only.
Power Requirements
100 to 240 VAC, 47 to 63 Hz, 30 Watts
Environmental
Requirements
32° to 122° F (0° to 50° C).
0% to 95% relative humidity (non-condensing).
Conformance Testing
International: EN60335-1:1994 Class 1
North America: UL 1310 Class 2
category
nominal specifications
Pads, Cable, and Connector
Disposable and self-adhesive. DP2/DP6 adult defibrillator pads have a nominal
active surface area of 100 cm
2
each and are provided in a sealed package with an
integrated 122 cm (48 inch), typical, cable and connector. M3870A FR2
infant/child reduced-energy defibrillator pads have a nominal active surface area of
44 cm
2
each and are provided in a sealed package with an integrated 122 cm (48
inch), typical, cable and connector incorporating attenuating electronics.