Maintaining, Testing, and Troubleshooting Your HeartStart FR2+

4-13

P

h

ili

ps

Me

dic

a

l

Sy

s

tem

s

4

Voice prompts:

ANALYZING

INTERRUPTED

or

CANNOT

ANALYZE

or

STOP

ALL

MOTION

• The patient is being moved or jostled.

• Radio or electrical sources are

interfering with ECG analysis.

• The environment is dry and movement

around the patient is causing static
electricity to interfere with ECG
analysis.

• Stop CPR; do not touch the

patient. Minimize patient motion. If
the patient is being transported,
stop the vehicle if needed.

• Check for possible causes of

radio and electrical interference
and remove them from the area.

• Responders and bystanders

should minimize motion,
particularly in dry environments
that can generate static electricity.

Voice and screen prompts:

NO

SHOCK

DELIVERED

The patient impedance is not appropriate
for the FR2+ to deliver a shock.

• Make sure the defibrillator pads

are correctly positioned on the
patient according to the diagram
on the back of the pads.

• Make sure the defibrillator pads

connector is completely inserted
in the connector socket.

• Press the defibrillator pads firmly

to the patient's chest.

• Replace the defibrillator pads if

necessary.

Voice prompt:

SHOCK

BUTTON

NOT

PRESSED

Shock has been advised but not delivered
within 30 seconds. (FR2+ has been
disarmed.)

• When next prompted, press the

Shock button to deliver shock.

symptom

possible cause

recommended action