beautypg.com

B technical specifications, Heartstart fr2+ defibrillator specifications – Philips M3860A User Manual

Page 69

background image

P

h

ili

ps

Me

dic

a

l

Sy

s

tem

s

B

B-1

B Technical

Specifications

The specifications for the HeartStart FR2+ provided in this chapter apply to
both the M3860A and M3861A, unless otherwise noted. Additional
information can be found in the

Technical Reference Manual

for the FR2

Defibrillator, located online at www.medical.philips.com.

HeartStart FR2+ Defibrillator Specifications

Physical

Environmental

category

nominal specifications

Size

2.6" high x 8.6" wide x 8.6" deep (6.6 cm x 21.8 cm x 21.8 cm).

Weight

Approximately 4.7 lbs (2.1 kg) with M3863A FR2 standard battery installed.
Approximately 4.5 lbs (2 kg) with optional M3848A FR2+ rechargeable battery
installed.

category

nominal specifications

Operating Temperature
and Humidity

32° to 122° F (0° to 50° C).
0% to 95% relative humidity (non-condensing).

Standby Temperature
and Humidity

32° to 109° F (0° to 43° C).
0% to 75% relative humidity (non-condensing).
Applies to HeartStart FR2+ with battery installed and stored with defibrillator
pads.

Altitude

Meets MIL-810E 500.3, Procedure II (-500 feet to 15,000 feet).

Shock/Drop Abuse
Tolerance

Meets MIL-STD-810E 516.4, Procedure IV (after a 1 meter drop to any edge,
corner, or surface, in standby mode).

Vibration

Meets MIL-STD-810E 514.4-17.

Sealing

With data card tray and battery installed, meets IEC 529 class IP54.

ESD

Meets EN 61000-4-2:1998 Severity Level 4.

EMI (Radiated)

Meets EN 60601-1-2 limits (1993), method EN 55011:1998 Group 1 Level B.

This manual is related to the following products: