NSA action
(minutes)
MONITOR
,
0.5, 1.0, 1.5,
2.0, 2.5, 3.0
MONITOR
Sets how the FR2+ behaves following a
NO
SHOCK
ADVISED
(NSA) decision:
MONITOR
— directs the FR2+ to monitor the patient’s
ECG following an NSA decision and to prompt the user
periodically to provide CPR. The interval for CPR
prompting is set by the Monitor Prompt Interval.
TIME
SETTING
— directs the FR2+ to provide a CPR
pause period following an NSA decision (NSA CPR
Pause).
• If no shocks have been delivered in the current Shock
Series (e.g., the patient’s initial monitored rhythm is
non-shockable), the length of the CPR pause is defined
by the NSA Action time setting.
• If shocks have been delivered in the current Shock
Series (e.g., the NSA decision follows a shock), the
length of the CPR pause is instead defined by the CPR
Timer setting.
CPR prompt
LONG
,
SHORT
LONG
Sets the level of detail provided in the CPR reminder voice
prompts provided at the completion of a Shock Series.
LONG
— provides detailed coaching to check airway,
breathing, and pulse before beginning CPR.
SHORT
— simply directs user to begin CPR if needed.
monitor prompt
interval
(minutes)
1.0, 1.5, 2.0,
2.5, 3.0,
∞
(infinite)
1.0
Sets the interval for patient care prompts provided during
FR2+ monitoring of the patient’s ECG following an NSA
decision. Selection of
∞
(infinite) means that no repeat
prompting will be provided during ECG monitoring.
parameter
settings
default
description