The highlighted top line is labeled

MANUAL

CHARGE

, with an arrowhead

pointing to the upper Option button.

If the ECG display shows that, in your expert clinical judgment, the patient
has a shockable rhythm, press the upper Option button (

MANUAL

CHARGE

).

The HeartStart FR2+ will immediately charge for shock delivery.

As soon as charging begins, the screen message changes to

CHARGING

,

STAND

CLEAR

, and the label for the arrowhead pointing to the upper Option

button changes to

MANUAL

DISARM

.

The FR2+ beeps while it is charging. When the beeping changes to a
continuous tone and the Shock button light flashes, press the Shock button
to deliver a shock. However, if the ECG display shows that the patient’s
rhythm has changed to a non-shockable rhythm, press the upper Option
button to disarm the HeartStart FR2+.

After shock delivery, the HeartStart FR2+ returns to the initial advanced
mode screen. To return to non-manual, AED mode operation, turn the FR2+
off by pressing the On/Off button. Then turn the FR2+ on by pressing the
On/Off button again.