System Functions Screen
System Functions
2
1
8
6
5
7
3
4
1
Sorter (Enabled/Disabled). If your imager is configured with no
sorter, this button will be absent. When touched, the sorter is enabled or
disabled. When disabled, the laser imager sends all film to bin 1.
2
Audible Alarm (Enabled/Disabled). When touched, the audible alarm
for error and status messages is enabled or disabled (Key Op function).
3
Energy Save Schedule. When touched, the Energy Save Schedule
screen is displayed for viewing (see
“Scheduling the Automatic Energy
). Editing the daily periods of Energy Save Mode
is a Key Op function.
4
More DVM Test Patterns. This button is present only if the
mammography upgrade is installed and enabled. When you touch this
button, a list of mammography test patterns appears. You can print these
test patterns to verify that your imager meets mammography quality
standards. See
“Printing a DVM (DryView Mammography) Test Pattern”
5
Density. When touched, Density test print is selected. When the Print
button is touched, the laser imager prints a Density test print from the
selected film drawer and at the selected density (default 3.00).