Safety and Related Information

Guidance and Manufacturer’s Declaration for Electromagnetic Immunity

The Kodak DryView 8900 Laser Imager is intended for use in the electromagnetic environment specified below. The
customer or user of the laser imager should assure that it is used in such an environment.

Immunity Test

IEC 60601 Test Level Compliance Level

Electromagnetic Environment -
Guidance

Electrostatic
discharge (ESD):

• EN61000-4-2
• IEC 61000-4-2

±

6 kV contact

±

8 kV air

±

6 kV contact

±

8 kV air

Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.

Electrical fast
transient/burst:

• EN61000-4-4
• IEC 61000-4-4

±

2 kV for power supply

lines

±

1 kV for input/output

lines

±

2 kV for power supply

lines

±

1 kV for input/output

lines

Mains power quality should be that of a
typical commercial or hospital
environment.

Surge:

• EN61000-4-5
• IEC 61000-4-5

±

1 kV differential mode

±

2 kV common mode

±

1 kV differential mode

±

2 kV common mode

Mains power quality should be that of a
typical commercial or hospital
environment.

Voltage dips, short
interruptions and
voltage variations
on power supply
lines:

• EN61000-4-11
• IEC 61000-4-11

<5% U

τ

* (>95% dip in

U

τ

*) for 0.5 cycle

40% U

τ

* (60% dip in

U

τ

*) for 5 cycles

70% U

τ

* (30% dip in

U

τ

*) for 25 cycles

<5% U

τ

* (>95% dip in

U

τ

*) for 5 sec.

<5% U

τ

* (>95% dip in

U

τ

*) for 0.5 cycle

40% U

τ

* (60% dip in

U

τ

*) for 5 cycles

70% U

τ

* (30% dip in

U

τ

*) for 25 cycles

<5% U

τ

* (>95% dip in

U

τ

*) for 5 sec.

Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Kodak
DryView
8900 Laser Imager requires
continued operation during power
mains interruptions, it is recommended
that the Kodak DryView 8900 Laser
Imager be powered from an
uninterruptible power supply or a
battery.

Power frequency
(50/60 Hz)
magnetic field:

• EN61000-4-8
• IEC 61000-4-8

3 A/m

3 A/m

Mains power quality should be that of a
typical commercial or hospital
environment.

NOTE: * U

τ

is the a.c. mains voltage prior to application of the test level.