Guidance and Manufacturer’s Declaration for Electromagnetic Immunity
The Kodak DryView 8900 Laser Imager is intended for use in the electromagnetic environment specified below. The
customer or user of the Kodak DryView 8900 Laser Imager should assure that it is used in such an environment.
Immunity
Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the Kodak DryView 8900 Laser
Imager, including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended Separation Distance:
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
Radiated RF
IEC 61000-4-3
3 v/m
80 MHz to 2.5 GHz
3 v/m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d is the recommended separation distance in meters (m).
P is the maximum output rating of the transmitter in watts (W)
according to the transmitter manufacturer.
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
a
, should be less than the
compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE:
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
See Note 1 on next page.
b
See Note 2 on next page.
d
1.17 P
=
d
1.17 P
=
d
2.33 P
=