54 inp:prob:coun, 55 inp:prob:del Page 95: 57 inp:prob:iden :prob:iden
Page 95: 57 inp:prob:iden :prob:iden
85
Remote Operation
Interface Commands
Adds a new probe definition to the Probe Library. Before a probe can be assigned to a channel and tem-
INP
ward slashes, and underscores. It must be between one and 24 characters in length, inclusive.
If
probe definition will be added and given the specified ID. The probe parameters will initially be set to default
values. The command INP:PROB:PAR should be used to set the parameters.
If
nition is added, and error -224, “Illegal parameter value” will be generated in the error queue.
8.7.3.54 INP:PROB:COUN?
Alternate parameter: MAX
Reads the number of probe definitions in the Probe Library.
8.7.3.55 INP:PROB:DEL
Example command: INP:PROB:DEL “PRT_A46002”
Password protection: Probe Protection
Deletes a probe definition from the Probe Library.
Parameter
Probe Library, nothing is deleted, and error -224, “Illegal parameter value” is generated in the error queue.
If the deleted probe was assigned to a channel the channel’s probe setting is changed to undefined (see Section
8.7.3.57, INP
8.7.3.56 INP:PROB:FIRS?
Example response: “PRT_A46002”
Reads the ID of the first probe in the Probe Library. It also resets the library index used by the
INP:PROB:NEXT? command.
If the Probe Library is empty the response is an empty string, “”.
8.7.3.57 INP
Example command: INP1:PROB:IDEN “PRT_A46002”
Example command: INP2:PROB:IDEN NONE
Query form: INP
Assigns a probe definition to the channel.
Suffix
for channels of the first attached scanner, and 15 through 24 for channels of the second attached scanner.
Parameter
closing quotations marks are optional. If
error -224,“Illegal parameter value” is generated in the error queue.
If a valid probe ID is assigned to the channel, the response to the query form is the ID of the probe enclosed in
quotation marks. If the assigned probe is undefined, the response is “NONE” without quotation marks.
The channel’s Calculation setting will automatically be set to resistance if it was temperature when the chan-
nel’s assigned probe is set to NONE.
8.7.3.58 INP:PROB:NEXT?
Example response: “A71040”