2 remote interface (f2), Remote interface (f2) – Fluke 1595A User Manual
Page 61
51
Menus and Screens
Main Menu
DAYLIGHT SAVING - determines whether the clock is automatically adjusted at the beginning and end of
the Daylight Saving period. Options are OFF, NORTH AMERICA, and EUROPE. The default option is OFF.
5.7.5.2 REMOTE INTERFACE (F2)
MAIN MENU | SYSTEM MENU |
The Remote Interface screen contains the function keys used for configuring and monitoring the remote inter-
Section 8, Remote Operation, on page 65.
The following are the function keys located in the Remote Interface menu.
5.7.5.2.1 SERIAL (F1)
MAIN MENU | SYSTEM MENU | REMOTE INTERFACE |
The Serial function key is used to configure the RS-232 serial port with the following configuration fields:
BAUD RATE - baud rate setting of the serial port. It may be set to 1200, 2400, 4800, 9600, 19200, or 38400.
The default setting is 9600.
LINEFEED - determines whether a linefeed character (ASCII code 10) is transmitted at the end of every
message.
5.7.5.2.2 USB (F2)
MAIN MENU | SYSTEM MENU | REMOTE INTERFACE |
The USB function key displays the following configuration field:
PORT ENABLE – enables and disables the rear USB port for remote communications with a computer (The
front USB port, used for an external USB memory device, is always enabled).
5.7.5.2.3 NETWORK (F3)
MAIN MENU | SYSTEM MENU | REMOTE INTERFACE |
The Network function key opens a screen allowing the user to configure the Network port settings.
The network port implements a 10/100 Base T Ethernet network interface. It supports the protocols TCP/IP,
DHCP, and HTML. The port is disabled by default and must be enabled to be used. The user may select DHCP
or specify a static network address.
The network port may be used for remote control of the Super-Thermometer by a PC and interface to network
software such as Telnet or a COM port redirector driver. In this mode, the network port accepts a raw TCP
connection on TCP port number 5025.
The network port also provides a simple HTML web page interface to the Super-Thermometer to allow remote
monitoring via a web browser.
The following are the configuration fields:
HARDWARE MAC - shows the hardware MAC address of the network port. It cannot be changed.
DHCP ENABLE - determines whether the static IP address is used or is assigned dynamically by a server.
The default option is ON (dynamic).
CAUTION:
When DHCP is enabled, do not remove the Ethernet cable and turn the power off. The cable
needs to remain plugged into the unit while cycling the power.
IP ADDRESS - specifies the network port’s IP address when DHCP is OFF, or shows the dynamically as-
signed IP address when DHCP is set to ON.
GATEWAY - specifies the gateway server’s IP address to be used when DHCP is OFF.,