Page 58: Field settings (f2)

1594A/1595A Super-Thermometer
Main Menu
48
When the User Settings screen appears after power-on, when the Start-Up Help setting is ON, the User Set-
tings screen includes function key NEXT (F1). This key leads to the Time and Date Settings screen (see Sec-
tion 5.7.5.1, TIME DATE (F1), on page 50), where the clock may be set to the local time.
5.7.4.2 FIELD SETTINGS (F2)
MAIN MENU | DISPLAY MENU |
The Field Settings function key is used for selecting statistical fields to display in the Measurement Screen. A
particular set of statistical fields can be stored as a set and recalled using the RECALL MEMORY selection.
Four different sets can be stored in Recall Memory, any of which can be restored to factory defaults using the
RESTORE DEFAULT (F1) function key (see Default Statistical Fields table below). Restore Default only af-
fects the Recall Memory selection that is currently display.
Individual statistical fields in any of the Recall Memory sets can be changed by first selecting the RECALL
MEMORY set number, pressing the ENTER key, arrowing down to the field to be changed, changing the field
with the Left/Right arrow keys and pressing the ENTER key save the change. The available data field options
and their descriptions are listed in the tables below.
Table 15 Default Statistical Fields
Field
Memory 1
Memory 2
Memory 3
Memory 4
1
MEAN
MEAN
MAXIMUM
MEAN
2
STD DEV
STD DEV
MINIMUM
STD DEV
3
SE OF MEAN
N
DELTA
TIME
4
RESISTANCE
RESISTANCE
RESISTANCE
DATE
Table 16 Statistical Field Descriptions
Option
Description
Equation
MEAN
arithmetic mean of measurements
Mean
n
y
i
n
i
=
=
∑
1
1
STD DEV
sample standard deviation of measurements
Std Dev
n
y
n Mean
i
i
n
=
−
− •
=
∑
1
1
2
1
2
SE OF MEAN
standard error (estimated standard deviation) of the
mean
SE of Mean
std dev
n
= (
)
MAXIMUM
maximum measurement
MINIMUM
minimum measurement
SPREAD
difference between maximum and minimum
Spread Maximum Minimum
=
−
N
number of measurements (sample size)
DELTA
difference relative to a delta channel or offset
RESISTANCE
resistance measurement
RATIO
resistance ratio measurement
TIME
time from the system clock
DATE
date from the system clock
As a note of further explanation, SE OF MEAN provides an estimate of the uncertainty (k = 1) of the MEAN
value. SE OF MEAN is calculated from the sample standard deviation of unfiltered measurements (not the
same as the displayed standard deviation of filtered measurements) and the sample size. The method used to
calculate SE OF MEAN disregards the effect of the digital filter since it causes correlation among displayed
measurements. Otherwise the result of the formula for SE OF MEAN would produce an erroneously low
value.