Interface Commands

72

Command

Function

Associated Setting or

Function (screen)

INIT:CONT

Starts or stops measurement

START/STOP MEASURING (Main
Screen)

INIT:STOP:BEEP

Sets the stop timer’s Timer Beep setting

Stop Timer Enable (Timing Settings)

INIT:STOP:DUR

Sets the stop timer’s Timer Duration setting

Stop Timer Duration (Timing
Settings)

INIT:STOP[:STAT]

Sets the Stop Timer Enable

Stop Timer Enable (Timing Settings)

INIT:STOP:TIME?

Reads the stop timer’s time remaining

Stop Timer time remaining (Main
Screen)

INP:CURR

Sets a channel’s Sensing Current setting

Sensing Current (Channel Settings)

INP:CURR:DATA?

Reads a channel’s actual sensing current

INP:CURR:STAN

Sets a channel’s Standby Current enable

Standby Current (Channel Settings)

INP:PROB:ADD

Adds a new probe definition to the Probe Library

ADD PROBE (Probe Menu)

INP:PROB:COUN?

Reads the number of probe definitions in the Probe
Library

INP:PROB:DEL

Deletes a probe definition from the Resistor Library

DELETE PROBE (Manage Probes)

INP:PROB:FIRS?

Reads the ID of the first probe in the Probe Library

INP:PROB:IDEN

Assigns a probe definition to the channel

ASSIGN PROBE (Channel Menu)

INP:PROB:NEXT?

Reads the ID of the next probe in the Probe Library

INP:PROB:PAR

Sets the value of a setting in a resistor definition

EDIT PROBE (Probe Menu)

INP:PROB:TEST?

Reads the temperature calculated from resistance
using the probe definition

TEST CALCULATION (Edit Probe)

INP:REAR:RS:IDEN

Sets the reference resistor assigned to a rear
reference resistor input

ASSIGN RESISTOR (Set Up Rs)

INP:RS:ADD

Adds a new resistor definition to the Resistor Library

ADD RESISTOR (Define Resistor)

INP:RS:COUN?

Reads the number of resistor definitions in the
Resistor Library

INP:RS:DEL

Delete the resistor ID from the Resistor Library

DELETE RESISTOR (Manage
Resistors)

INP:RS:FIRS?

Reads the ID of the first resistor in the Resistor
Library

INP:RS:IDEN

Sets the reference resistor assigned to a front
reference resistor input

ASSIGN RESISTOR (Set Up Rs)

INP:RS:NEXT?

Reads the ID of the next resistor in the Resistor
Library

INP:RS:OVEN:STAB

Reads the oven stable condition

INP:RS:PAR

Sets the value of a setting in a resistor definition

EDIT RESISTOR (Define Resistor)

INP:RS:ROUT

Sets a channel’s Reference Resistor (Rs) setting

ROUT:CLOS

Selects one channel for measurement

ROUT:CLOS:STAT?

Returns the channel number of whatever channel is
actively being measured

Active Channel (Main Screen)

ROUT:SCAN

Enables channels for scanning

ROUT:SCAN:ADD

Adds channels to the scan list

ROUT:SCAN:ALT

Sets the scan Alternate Reference setting

Alternate Reference (Scan Settings)

ROUT:SCAN:DEL

Removes channels from the scan list

ROUT:SCAN:REF

Sets the scan Reference Channel setting

Reference Channel (Scan Settings)

ROUT:SCAN:REP

Sets the scan Repeat Count setting

Repeat Count (Scan Settings)

ROUT:SCAN:STAT

Sets the Scan Enable setting

Scan Enable (Scan Settings)

SENS:APER

Sets the measurement time and Display Interval

Display Interval (Timing Settings)