71
Remote Operation
Interface Commands
Command
Function
Associated Setting or
Function (screen)
CAL:DATE
Sets the instrument’s date of calibration
Calibration Date (Calibration
Parameters)
CAL:PAR:LIN
Sets a calibration linearity correction parameter
Linearity C
Parameters)
CAL:PAR:RES
Sets the calibrated resistance of an internal
reference resistor
—
W Resistor (Calibration
Parameters)
CAL:PAR:ZERO
Sets the calibration zero correction parameter
Zero Correction (Calibration
Parameters)
CALC
Reads a channel’s statistical values
(Main Screen)
CALC
Reads a channel’s statistical mean
Mean (Main Screen)
CALC
Clears a channel’s statistics
RESET STATISTICS (Main Screen)
CALC:AVER:COUN
Sets the statistics Sample Count setting
Sample Count (Statistics Settings)
CALC
Reads a channel’s statistical maximum
Maximum (Main Screen)
CALC
Reads a channel’s statistical minimum
Minimum (Main Screen)
CALC
Reads a channel’s statistical sample standard
deviation
Std Dev (Main Screen)
CALC
Reads a channel’s statistical standard error of the
mean
SE of Mean (Main Screen)
CALC
Reads a channel’s statistical spread
Spread (Main Screen)
CALC
Reads a channel’s statistical sample size
N (Main Screen)
CALC:AVER:TCON
Sets the Statistics Mode setting
Statistics Mode (Statistics Settings)
CALC:DELT:CHAN
Sets the Delta Channel setting
Delta Channel (Statistics Settings)
CALC
Reads a channel’s delta result
Delta (Main Screen)
CALC:DELT:MODE
Sets the Delta Mode setting
Delta Mode (Statistics Settings)
CALC:DELT:OFFS
Sets the Delta Offset setting
Delta Offset (Statistics Settings)
CALC
Sets the channel’s Calculation setting
Calculation (Channel Settings)
DATA:REC:ALL
Sets the data recording Record Statistics setting
Record Statistics (Recording
Settings)
DATA:REC:FILE
Sets the data recording USB File setting
USB File (Recording Settings)
DATA:REC:MEM
Sets the data recording Memory Device setting
Memory Device (Recording Settings)
DATA:REC[:STAT]
Enables data recording
START/STOP RECORDING (Main
Screen)
DISP:ALER:CLEA?
Clears any alert message that is active
CLEAR MESSAGE (Main Screen)
DISP:ALER[:READ]?
Reads the display alert message
(Main Screen)
DISP:DATE:FORM
Sets the display Date Format setting
Date Format (Display Settings)
DISP:DEC:FORM
Sets the display Decimal setting
Decimal (Display Settings)
DISP:FIEL:FEED
Reads a statistics field setting of the visible statistics
fields memory set
(Main Screen)
DISP:FIEL:MEM[?]
Reads or sets the statistics fields memory set for
display
Recall Memory (Field Settings)
DISP:FIEL:MEM
Reads or sets a statistics field setting of a statistics
fields memory set
Field
DISP:LANG
Sets the display Language setting
Language (Display Settings)
DISP:TEMP:DIG
Sets the display Temperature Resolution setting
Temperature Resolution (Display
Settings)
DISP:WARN:DUE
Sets the display Rs Due Warning enable
Rs Due Warning (Display Settings)
DISP:WARN:ITS
Sets the display ITS-90 Subrange Warning enable
ITS-90 Subrange Warning (Display
Settings)
FETC?