6 sense:resistance:range |min|max|def 92, 7 sense:resistance:wire? [min|max|def] 92, Sense:resistance:range |min|max|def – Fluke 1529 Chub-E4 User Manual
Page 98: Sense:resistance:wire? [min|max|def

0
This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
The SENSE suffix,
to 14). Specifying the MIN parameter returns a value of 0. Specifying the MAX
parameter returns a value of 1. Specifying the DEF parameter returns a value of
0. The *RST command sets the resistance range to the default (DEF).
8.4.2.6
SENSe
Sets the resistance range setting.
SENS4:RES:RANG MIN
This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
The SENSE suffix,
to 14). The
the MIN, MAX or DEF parameter sets the resistance range setting to the mini-
mum, maximum, or default value respectively.
This command may be password-protected. To access this command:
• If the SYST:PASS:CONV? command returns a 0 (probe password-protec-
tion is disabled), the password is not required.
• If the SYST:PASS:CONV? command returns a 1 (probe password-protec-
tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN
mands).
8.4.2.7
SENSe
Returns the probe wire setting.
SENS1:RES:WIRE?
3
This command only applies to PRT/Thermistor channels. Executing this com-
mand on Thermocouple channels generates a -294, "Incompatible type", error
message.
The SENSE suffix,
to 14). Specifying the MIN parameter returns a value of 2. Specifying the
MAX parameter returns a value of 4. Specifying the DEF parameter returns a
value of 4.
1529 Chub-E4 Thermometer Readout
User’s Guide
92