6 route:scan? 95, 7 route:scan 95, 8 route:scan:mode? [min|max|def] 95 – Fluke 1529 Chub-E4 User Manual
Page 101: 9 route:scan:mode |min|max|def 95, 4 probe commands 95, 1 calculate:convert:catalog? 95, Route:scan, Route:scan |min|max|def

8.4.3.6
ROUTe:SCAN?
Returns a comma delimited list of the enabled channels.
ROUT:SCAN?
(@2,3,4)
The list of channels is enclosed in parenthesis and preceded with the @
symbol.
8.4.3.7
ROUTe:SCAN
Sets the channels for scanning.
ROUT:SCAN 2,3
The
enable. Channels that are omitted from this list are disabled. If the
parameter is omitted, all channels are disabled. The *RST command sets chan-
nel 1 ON, and all other channels OFF.
8.4.3.8
ROUTe:SCAN:MODE? [MIN|MAX|DEF]
Returns the current scan mode setting.
ROUT:SCAN:MODE? MIN
0
If 0 is returned, the scan mode is simultaneous. If a 1 is returned the scan mode
is scan. Specifying the MIN parameter returns a value of 0. Specifying the
MAX parameter returns a value of 1. Specifying the DEF parameter returns a
value of 1.
8.4.3.9
ROUTe:SCAN:MODE
Sets the scan mode setting.
ROUT:SCAN:MODE 1
The
ifying the MIN, MAX, or DEF parameter sets the mode to the minimum, maxi-
mum, or default, respectively. The *RST command sets scan mode to scan.
8.4.4
Probe Commands
The probe commands are used for querying, setting, and copying the probe
conversion type, conversion parameters, and serial number and for testing the
temperature conversion for a channel. The probe commands for copying and
setting parameters are password protected.
8.4.4.1
CALCulate
Returns a list of conversion types available for the specified channel.
95
8 Digital Communications Interface
Commands