beautypg.com

2 calculate:convert:copy |all 96, 3 calculate:convert:name? 96, 4 calculate:convert:name 96 – Fluke 1529 Chub-E4 User Manual

Page 102: Calculate:convert:copy |all, Calculate:convert:name, Calculate:convert:name

background image

CALC2:CONV:CAT?

“RES”,“ITS”,”ITS5”,”PT”,”CVD”,”TRES”,”TTEM”,”YSI”

The CALCulate suffix, , specifies the channel (1 to 4) or memory chan-
nel (5 to 14). Each item in the response list is enclosed in double quotes and
separated from other items by a comma. See Table 7 on page 128, for the avail-
able conversion types.

8.4.4.2

CALCulate:CONVert:COPY |ALL

Copies the conversion type, sub-ranges (ITS-90), serial number, and character-
ization coefficients from one channel to another.

CALC2:CONV:COPY 8

The CALCulate suffix specifies the source channel (1 to 4). The chn> parameter specifies the destination channel (1 to 4). When the copy is
performed, the source channel parameters are not affected. If the destination
channel type is different from the source channel type, the error -294, "Incom-
patible type" is generated. Specifying the ALL parameter copies the source
channel parameters to all compatible channels. The example above copies the
parameters from channel 2 to memory channel PMEM3.

This command may be password-protected. To access this command:

• If the SYST:PASS:CONV? command returns a 0 (probe password-protec-

tion is disabled), the password is not required.

• If the SYST:PASS:CONV? command returns a 1 (probe password-protec-

tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN command (see Section 9.4.11, Password Com-
mands).

8.4.4.3

CALCulate:CONVert:NAMe?

Returns the name of the selected conversion type for the specified channel.

CALC2:CONV:NAME?

ITS

The CALCulate suffix, , specifies the channel (1 to 4) or memory chan-
nel (5 to 14). The conversion types and their names are listed in Table 7 on
page 128.

8.4.4.4

CALCulate:CONVert:NAMe

Sets the conversion type for the specified channel.

CALC2:CONV:NAME CVD

The CALCulate suffix, , specifies the channel (1 to 4) or memory chan-
nel (5 to 14). The parameter is a mnemonic indicating the conversion

1529 Chub-E4 Thermometer Readout
User’s Guide