2 calculate:convert:copy |all 96, 3 calculate:convert:name? 96, 4 calculate:convert:name 96 – Fluke 1529 Chub-E4 User Manual
Page 102: Calculate:convert:copy |all, Calculate:convert:name, Calculate:convert:name

CALC2:CONV:CAT?
“RES”,“ITS”,”ITS5”,”PT”,”CVD”,”TRES”,”TTEM”,”YSI”
The CALCulate suffix,
nel (5 to 14). Each item in the response list is enclosed in double quotes and
separated from other items by a comma. See Table 7 on page 128, for the avail-
able conversion types.
8.4.4.2
CALCulate
Copies the conversion type, sub-ranges (ITS-90), serial number, and character-
ization coefficients from one channel to another.
CALC2:CONV:COPY 8
The CALCulate suffix
performed, the source channel parameters are not affected. If the destination
channel type is different from the source channel type, the error -294, "Incom-
patible type" is generated. Specifying the ALL parameter copies the source
channel parameters to all compatible channels. The example above copies the
parameters from channel 2 to memory channel PMEM3.
This command may be password-protected. To access this command:
• If the SYST:PASS:CONV? command returns a 0 (probe password-protec-
tion is disabled), the password is not required.
• If the SYST:PASS:CONV? command returns a 1 (probe password-protec-
tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN
mands).
8.4.4.3
CALCulate
Returns the name of the selected conversion type for the specified channel.
CALC2:CONV:NAME?
ITS
The CALCulate suffix,
nel (5 to 14). The conversion types and their names are listed in Table 7 on
page 128.
8.4.4.4
CALCulate
Sets the conversion type for the specified channel.
CALC2:CONV:NAME CVD
The CALCulate suffix,
nel (5 to 14). The
1529 Chub-E4 Thermometer Readout
User’s Guide
96