20 logging:demand:store 111, 21 logging:demand:value? |min|max|def 111, 22 logging:label:name? 111 – Fluke 1529 Chub-E4 User Manual
Page 117: 23 logging:label:name

8.4.7.20
LOGging:DEMand:STORe
Stores the current measurement into the demand log.
LOG:DEM:STOR
If simultaneous scan mode is enabled, one reading is stored in the log for each
enabled channel.
8.4.7.21
LOGging:DEMand:VALue?
Returns the specified demand log entry.
LOG:DEM:VAL? 10
DATA_25,2,9.961,KO,14,59,35,2000,9,5
The
ther the MIN or DEF parameter returns the first entry in the demand log. Spec-
ifying the MAX parameter returns the last entry in the demand log. Each entry
in the demand log consist of the label, channel number, measurement value,
scale, time and date.
8.4.7.22
LOGging:LABel
Returns the name of the specified label.
LOG:LAB18:NAME?
DATA_18
The label suffix,
The label name is returned in string format.
8.4.7.23
LOGging:LABel
Sets the name of the specified label.
LOG:LAB18:NAME BATH
The label suffix,
names can be up to 8 characters in length and can include any letters, numeric
digits, and the underscore (‘_’).
8.4.8
System Commands
The system commands are used for querying instrument identification parame-
ters, temperature units, and for setting operating parameters to the defined de-
fault values.
8.4.8.1
*IDN?
Returns the instrument identification string that indicates the manufacturer,
model number, serial number, and code firmware version.
111
8 Digital Communications Interface
Commands