8 calculate:convert:print [|all [,]] 98, 9 calculate:convert:snumber? 98, 10 calculate:convert:snumber 98 – Fluke 1529 Chub-E4 User Manual
Page 104: Calculate:convert:print [|all [,:convert:snumber, Calculate:convert:snumber

CALC2:CONV:PAR:VAL RTPW, 100.0145, A4, 0.0045
The CALCulate suffix,
nel (5 to 14). The parameter is the name of the parameter as given in
Table 8, Conversion Types and Parameters. If the parameter is not ap-
plicable to the current conversion type, an error -221, “Settings Conflict”, is
generated. The
parameters can be set using a single command by separating the and
the value must immediately follow the parameter name.
This command may be password-protected. To access this command:
• If the SYST:PASS:CONV? command returns a 0 (probe password-protec-
tion is disabled), the password is not required.
• If the SYST:PASS:CONV? command returns a 1 (probe password-protec-
tion is enabled), the password must be successfully entered using the
SYST:PASS:CEN
mands).
8.4.4.8
CALCulate:CONVert:PRINt [
Prints the conversion parameters for the specified channel.
CALC:CONV:PRIN 2
“PRB”,CHAN2,”ser”,090105,”TYP”,RES,”RANGE”,0
If no parameters are specified, the ALL parameter is assumed and the parame-
ters for all channels are printed to the current port. Specifying the
rameter prints only the parameters for the specified channel (1 to 4) to the
current port. Optionally, the parameters may be printed to a specific port by
specifying the
8.4.4.9
CALCulate
Returns the probe serial number for the specified channel.
CALC2:CONV:SNUM?
A_336C
The CALCulate suffix,
nel (5 to 14). The returned serial number is in string format.
8.4.4.10
CALCulate
Sets the probe serial number for the specified channel.
CALC2:CONV:SNUM A_336C
The CALCulate suffix,
nel (5 to 14). The
1529 Chub-E4 Thermometer Readout
User’s Guide
98