The WINDow suffix,
two numbers separated by a comma. The first number represents the channel
associated with the field. The second number represents the parameter to be
displayed in the specified field. See Table 5 on page 126, Field Types, for the
field types.
8.4.6.10
DISPlay:WINDow
Sets the specified field within the specified display set.
DISP:WIND1:FIEL3:FEED 0,10
The WINDow suffix,
specifies channel (1 to 4) associated with the field. The
ifies the parameter to be displayed in the specified field. See Table 5 on page
126, Field Types, for the field types.
8.4.7
Logging Commands
The logging commands are used for logging data automatically and on demand.
8.4.7.1
LOGging:AUTomatic:DELete [
Deletes entries from the auto log.
LOG:AUT:DEL 10
Specifying a number in the
that label number (1 to 25) to be deleted. If the ALL parameter is specified or
no parameter is specified, all auto log entries are deleted.
8.4.7.2
LOGging:AUTomatic:COUNt? [MIN|MAX|DEF]
Returns the number of measurements to be stored in the auto log during an auto
log session.
LOG:AUT:COUN?
5000
Specifying the MIN parameter returns a value of 1. Specifying the MAX pa-
rameter value returns a value of 8160. Specifying the DEF parameter returns a
value of 8160.
8.4.7.3
LOGging:AUTomatic:COUNt
Sets the number of measurements to be stored to the auto log during an auto
log session.
LOG:AUT:COUN 100
1529 Chub-E4 Thermometer Readout
User’s Guide
106