The A24 Registers

The following A24 registers are accessible at the base address defined by
the device’s offset Register. The registers at offsets 0 to E

$

are not

accessible using longword (D32) accesses. The registers at offsets 10

$

to

FFFFF

$

may be accessed by any of the of the D08(EO), D16, or D32

modes.

FFFFF

$

8000 0

$

Movable DSP

Bus Window

7FFFF

$

3000 0

$

Fixed DSP

Bus Window

2FFFF

$

2000 0

$

Send/Receive

Data Registers

1FFFF

$

0004 F

$

Fixed DSP

Bus Window

0003 F

$

0000 0

$

VXI Bus A16

Registers

The A24 registers are defined as follows:

q

VXI Bus A16 Registers: These are the same registers accessed at the device’s
A16 base address.

q

Fixed DSP Bus Window: Accesses to this region are mapped to the
corresponding locations at the base of the internal DSP’s memory map, also
accessible through Page 0 of the moveable DSP bus window.

q

Send/Receive Data Registers: Accesses to any address in this region will
read/write the Send and Receive Data registers defined in the A16 register set.
VME Bus D32 Block Transfers are supported for these addresses only.

q

Movable DSP Bus Window: Accesses to this region are mapped (by the Page
Map register) to different 512 kB regions of the internal DSP bus.

HP E1432A User's Guide
Register Definitions

A-4