A.4

Setting Range Values (Local Zero and Span),

Continued

Procedure, continued

Table A-3

Setting Range Values Using Local Zero and Span

Adjustments,

Continued

Step Action

10

Wait 30 seconds so that changes have been copied to the
transmitter’s non-volatile memory.

11

Remove applied pressure and turn OFF transmitter power.

12

Replace end-cap on PWA side of electronics housing and tighten
lock.

13

Remove milliammeter from TEST terminals and replace end-cap and
tighten lock.

14

Turn ON transmitter power and check smart meter reading, if
applicable.

Figure A-2

Typical Setup for Setting Range Values Using Local Zero and Span Adjustments.