SAFETY INFORMATION

ESD declaration statement

There was flickered disturbance on the screen during the test, but could self-recover after the test.

(This permissive loss of performance was met the specification of EUT.)

DIP declaration statement

The EUT reset during the test, but could self-recover after the test.

(This permissive loss of performance was met the specification of EUT.)

Recommended separation distances between portable and mobile RF communications equipment

and the DR-17G and DR-22G

The DR-17G and DR-22G is intended for use in an electromagnetic environment in which radiated RF

disturbances are controlled. The customer or the user of the DR-17G and DR-22G can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF

communications equipment (transmitters) and the DR-17G and DR-22G as recommended below, according
to the maximum output power of the communications equipment.

Rated maximum output

power of transmitter (W)

Separation distance according to frequency of transmitter (m)

150kHz to 80MHz

1.2

80MHz to 800MHz

1.2

800MHz to 2.5GHz

2.

0.01

0.12

0.12

0.23

0.1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

For transmitters rated at a maximum output power not listed above, the recommended separation distance

“d” in meters (m) can be estimated using the equation applicable to the frequency of the transmitter,

where “P” is the maximum output power rating of the transmitter in watts (W) according to the transmitter

manufacturer.

NOTE 1

At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2

These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

Cable length

Power Cord : Accessory

1.8m