SAFETY INFORMATION

Guidance and manufacturer’s declaration – electromagnetic immunity

The

DR-17G and DR-22G is intended for use in the electromagnetic environment specified below. The

customer or the user of the DR-17G and DR-22G should assure that it is used in such an environment.

Not Life-supporting Medical Equipment.

Immunity test

IEC/EN60601

test level

Compliance level

Electromagnetic environment –

guidance

Conducted RF

IEC/EN61000-4-6

Radiated RF

IEC/EN61000-4-3

3Vrms

150kHz to 80MHz

3V/m

80MHz to 2.5GHz

3Vrms 3V/m

Portable and mobile RF

communications equipment should

be used no closer to any part of the

DR-17G and DR-22G, including

cables, than the recommended

separation distance calculated

from the equation applicable to the

frequency of the transmitter.

Recommended Separation distance

d = 1.2 √ P d = 1.2 √ P, 80MHz to
800MHz
d = 2.3 √ P, 800MHz to 2.5GHz

Where “P”is the maximum output

power rating of the transmitter in

watts (W) according to the transmitter

manufacturer and “d” is the

recommended separation distance in

meters (m).

Field strengths from fixed RF

transmitters, as determined by an

electromagnetic site survey:

a. should be less than the

compliance level in each

frequency range

b. Interference may occur in the

vicinity of equipment marked with

the following symbol

NOTE 1

At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2

These guidelines may not apply in all situations. Electromagnetic propagation is affected by

absorption and reflection from structures, objects and people.

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and

land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted

theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the

DR-17G and DR-22G is used exceeds the applicable RF compliance level above, the DR-17G and DR-22G

should be observed to verify normal operation. If abnormal performance is observed, additional measures

may be necessary, such as reorienting or relocating the DR-17G and DR-22G.