SAFETY INFORMATION

Guidance and manufacturer’s declaration – electromagnetic immunity

The

DR-17G and DR-22G is intended for use in the electromagnetic environment specified below. The

customer or the user of the DR-17G and DR-22G should assure that it is used in such an environment.

Not Life-supporting Medical Equipment.

Immunity test

IEC/EN60601

test level

Compliance level

Electromagnetic environment –

guidance

Electrostatic

discharge (ESD)

IEC/EN61000-4-2

±8 kV contact
±15 kV air

±8 kV contact
±15 kV air

Floors should be wood, concrete

or ceramic tile. If floors are covered

with synthetic material, the relative

humidity should be at least 30%.

Electrical fast

transient / burst

IEC/EN61000-4-4

±2 kV for power supply

lines

±1 kV for input/output

lines

±2 kV for power supply

lines

±1 kV for input/output

lines

Mains power quality should be that

of a typical commercial or hospital

environment.

Surge

IEC/EN61000-4-5

For power supply lines:

± 1 kV line(s) to line(s)
± 2 kV line(s) to earth

For outdoor signal lines:

± 2 kV line(s) to earth

For power supply lines:

± 1 kV line(s) to line(s)
± 2 kV line(s) to earth

For outdoor signal lines:

± 2 kV line(s) to earth

Mains power quality should be that

of a typical commercial or hospital

environment.

Voltage dips, short

interruptions and

voltage variations

on power supply

input lines

IEC 61000-4-11

0% UT; 0,5 cycle

0% UT; 1 cycle

70% UT; 25/30 cycles

<5% UT; 250/300 cycles

0% UT; 0,5 cycle

0% UT; 1 cycle

70% UT; 25/30 cycles

<5% UT; 250/300 cycles

Mains power quality should be that

of a typical commercial or hospital

environment. If the user of the

DR-17G and DR-22G requires

continued operation during

power mains interruptions, it is

recommended that the DR-17G

and DR-22G be powered from an

uninterruptible power supply or a

battery.

Power frequency

(50/60Hz)
magnetic field

IEC/EN61000-4-8

3A/m

3A/m

Power frequency magnetic

fields should be at levels

characteristic of a typical location

in a typical commercial or hospital

environment.

NOTE:

UT is the a.c. mains voltage prior to application of the test level.

Mains power quality should be that of a typical commercial or hospital environment. If the user of the DR-17G

and DR-22G requires continued operation during power mains interruptions, it is recommended that the

DR-17G and DR-22G be powered from an uninterruptible power supply or a battery.