Safety information, Emc information – AG Neovo DR-22G 22" Full HD Dental Monitor (White) User Manual
Page 6

6
SAFETY INFORMATION
EMC Information
CAUTION
The DR-17G and DR-22G requires special precautions regarding EMC and need to be installed, put into
service and used according to the following information.
Do not use any cables other than the cables that provided or specified by us. Using other cables may cause
the increase of emission or decrease of immunity.
Do not put any portable and mobile RF communications equipment close to the DR-17G and DR-22G. Doing
so may affect the DR-17G and DR-22G.
The DR-17G and DR-22G should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation in the
configuration in which it will be used.
Anyone who connects additional equipment to the signal input part or signal output parts, configuring a
medical system, responsible that the system complies with the requirements of IEC/ EN60601-1-2.
Guidance and manufacturer’s declaration – electromagnetic emissions
The DR-17G and DR-22G is intended for use in the electromagnetic environment specified below. The
customer or the user of the DR-17G and DR-22G should assure that it is used in such an environment.
Not Life-supporting Medical Equipment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR11/EN55011
Group 1
The DR-17G and DR-22G uses RF energy only for its
internal function. Therefore, its RF emission are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR11/EN55011
Class B
The DR-17G and DR-22G is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC/EN61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC/EN61000-3-3
Class A