SAFETY INFORMATION

EMC Information

CAUTION

The DR-17G and DR-22G requires special precautions regarding EMC and need to be installed, put into

service and used according to the following information.

Do not use any cables other than the cables that provided or specified by us. Using other cables may cause

the increase of emission or decrease of immunity.

Do not put any portable and mobile RF communications equipment close to the DR-17G and DR-22G. Doing

so may affect the DR-17G and DR-22G.

The DR-17G and DR-22G should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation in the
configuration in which it will be used.
Anyone who connects additional equipment to the signal input part or signal output parts, configuring a

medical system, responsible that the system complies with the requirements of IEC/ EN60601-1-2.

Guidance and manufacturer’s declaration – electromagnetic emissions

The DR-17G and DR-22G is intended for use in the electromagnetic environment specified below. The

customer or the user of the DR-17G and DR-22G should assure that it is used in such an environment.

Not Life-supporting Medical Equipment.

Emissions test

Compliance

Electromagnetic environment – guidance

RF emissions

CISPR11/EN55011

Group 1

The DR-17G and DR-22G uses RF energy only for its

internal function. Therefore, its RF emission are very

low and are not likely to cause any interference in

nearby electronic equipment.

RF emissions

CISPR11/EN55011

Class B

The DR-17G and DR-22G is suitable for use in all

establishments, including domestic establishments

and those directly connected to the public low-voltage

power supply network that supplies buildings used for

domestic purposes.

Harmonic emissions

IEC/EN61000-3-2

Class A

Voltage fluctuations /
flicker emissions

IEC/EN61000-3-3

Class A