Aoutpr source o-rang u-rang break nmax, Pd865 snooper modbus, Serial input meter instruction manual – Precision Digital PD865 User Manual
Page 56
PD865 Snooper Modbus
Serial Input Meter
Instruction Manual
56
Analog Output Programming (aoutpr)
The 4-20 mA analog output can be programmed for its source of data,
overrange and underrange, absolute max and min output, and
communications break values. It can also be calibrated manually or
automatically.
Aoutpr
sourCe
O-rang
U-rang
break
nmax
Press Enter/Ack to
Make Selections
Press Up Arrow to
Scroll Through Choices
Press Menu to Exit at
any Time
disp
nmax
nmin
nmin
reg
Calib
nman
auto
4 nma
20 nma
Enter Actual
4 mA Value
Enter Actual
20 mA Value
Enter Over-range
Value
Enter Under-range
Value
Enter
Communication-
Break Value
Enter Minimum
Output Value
Enter Maximum
Output Value
*
Communication-Break menu
available only in Master Mode.
*
Factory Use
Only!
The overrange and underrange values are the values that will be
output when the display shows an overrange or underrange
condition.
The max and min values are the absolute limits for the 4-20 mA
output.
The communications break value determines the mA output when a
slave fails to reply to a command within the Response time.
For calibration instructions, see 4-20 mA Output Calibration,on
page 57.