beautypg.com

Power connection, Mounting, Power connection / mounting – HEIDENHAIN ND 930 User Manual

Page 35

background image

35

Power Connection / Mounting

To mount the display unit on a support, use the M4 threaded holes in
the rubber feet. You can also mount the display unit on the optional
tilting base.

Mounting

Hot leads: and
Protective ground:

Danger of electrical shock!
Connect a protective ground. This connection must
never be interrupted.

Unplug the power cord before opening the housing.

To increase the noise immunity, connect the ground terminal
on the rear panel to the central ground point of the machine.
(Minimum cross-section: 6 mm

2

)

The display unit will operate over a voltage range of 100 V to 240 V AC.
A voltage selector is not necessary.

Danger to internal components!
Use only original replacement fuses.
Two line fuses and a fuse for the switching outputs are inside
the housing.
Fuse types:

Line: F 2.5 A 250 V
Switching outputs: F 1 A

L

N

Power Connection

HE

IDE

NH

AIN

Tilting base

Support

This manual is related to the following products: