beautypg.com

Service, Power-up diagnostics failure, Troubleshooting – Grass Valley 2031RDA-MM User Manual

Page 33

background image

2031RDA-SM/-MM Instruction Manual

33

Service

Service

The 2031RDA-SM/-MM modules make extensive use of surface-mount
technology and programmed parts to achieve compact size and adherence
to demanding technical specifications. Circuit modules should not be ser-
viced in the field unless otherwise directed by Customer Service.

Power-up Diagnostics Failure

If the module has not passed self-diagnostics, do not attempt to trouble-
shoot. Return the unit to Grass Valley (see

Module Repair

).

Troubleshooting

If your module is not operating correctly, proceed as follows:

Check frame and module power. If power is not present, check the fuse
and voltage testpoints on the module as illustrated in

Figure 16

.

Check for presence and quality of input signals.

Verify that source equipment is operating correctly.

Check cable connections.

Figure 16. Location of Module Fuse

8318_03

F1

2A

125V

J108

J108

Voltage

Test Points

Fuse: F1, 2A, 125V

+24V

+5V

GND

GND

+3.3V

TP 6

GND