beautypg.com

Sloan 110 Dolphin Flushometer User Manual

Page 5

background image

I. Flushometer does not function (no flush).
A. Control Stop or Main Valve is Closed. Open Control Stop or Main Valve.

B. Handle Socket Assembly is damaged. Replace Handle Socket Assembly

(DO-20).

2. Volume of water is not sufficient to siphon fixture.
A. Control Stop is not open enough. Adjust Control Stop for desired delivery of

water volume.

B. Timing of Flushometer closure is set too fast. Remove DO-1 Cover Nut.

Use a small blade screwdriver to turn DO-3 Regulating Screw clockwise.

Flushometer should be set so that valve closure occurs between 7 and 12

seconds.

C. Water supply volume or pressure is inadequate. If no gauges are available

to properly measure supply pressure or volume of water at the Valve, then

completely open the Control Stop and hold down the Flushometer Handle

without allowing it to close.

• If the supply is adequate to siphon the fixture, increase timing of

closure. Remove DO-1 Cover Nut. Using small blade screwdriver,

turn DO-3 Regulating Screw clockwise until a 7 to 12 second flush is

achieved.

• If the supply is

NOT adequate to siphon the fixture, steps should be

taken to increase the supply pressure and/or volume. Contact the fixture

manufacturer for minimum water supply requirements of the fixture.

3. Flushometer closes immediately.
A. Oil chamber of upper body is empty. Remove DO-1 Cover Nut. Fill oil

chamber with SAE 10 Oil.

B. Piston seal is worn or damaged. Replace DO-6 Piston O-ring seal. Refer to

Dolphin Maintenance Guide for complete instructions.

4. Length of flush is too short (short flush).
A. Volume of oil in oil chamber of upper body is low. Remove DO-1 Cover Nut.

Fill oil chamber with SAE 10 Oil.

B. Timing of Flushometer closure is not adequately adjusted. Remove DO-1

Cover Nut. Using small blade screwdriver, turn DO-3 Regulating Screw

clockwise until a 7 to 12 second flush is achieved.

5. Length of flush is too long (long flush).
A. Timing of Flushometer closure is not adequately adjusted. Remove DO-1

Cover Nut. Using small blade screwdriver, turn DO-3 Regulating Screw

COUNTERCLOCKWISE until a 7 to 12 second flush is achieved.

6. Oil is leaking from handle.
A. Stem seals are worn or damaged. Replace two (2) SH-75 Throttle Seal

O-rings.

7. Valve does not close completely (water trickle).
A. Compression Spring is worn or damaged. Replace DO-16 Compression

Spring. Refer to Dolphin Maintenance Guide for complete instructions.

B. Flex-Sac Assembly is worn. Replace DO-15 Flex-Sac. Refer to Dolphin

Maintenance Guide for complete instructions.

When assistance is required, please contact the Sloan Technical Support at:

1-888-SLOAN-14 (1-888-756-2614)

5

A

Make sure Control Stop is CLOSED. Turn on water supply.

B

Open Control Stop. Activate Dolphin Flushometer Valve by pushing

Handle down. Hold the Handle down until all dirt and debris is

flushed out from the supply line and the water flowing into the

fixture is clean.

D

Open Control Stop COUNTERCLOCKWISE one FULL turn from

closed position. Activate Flushometer.

E

Adjust Control Stop after each flush until the rate of flow delivered

properly cleanses the fixture.

Release the Handle to begin the flush cycle. When the Handle

returns to the UP position, the flush cycle will finish.

C

The Sloan Dolphin Flushometer is engineered for quiet operation. Excessive water

flow creates noise, while too little water flow may not satisfy the needs of the fixture.

Proper adjustment is made when the plumbing fixture is cleansed after each flush

without splashing water out from the lip AND a quiet flushing cycle is achieved.
Never open Control Stop to where the flow from the valve exceeds the flow

capability of the fixture. In the event of a valve failure, the fixture must be able to

accommodate a continuous flow from the valve.

!!! IMPORTANT !!!

A

Remove the Cover Nut from the Valve body. Using a

flat blade screwdriver, slightly rotate the Regulating

Screw located in the center of the oil chamber.

F

After Control Stop adjustment is complete, install Stop Cap onto

the Control Stop. Wrench tighten Stop Cap to prevent vandalism.

The Sloan Dolphin Flushometer can be adjusted to meet the flushing requirements of any plumbing fixture. The flush volume, in gallons per flush (gpf) or liters

per flush (Lpf), is adjusted by turning the Regulating Screw. Flush volumes will vary with pressure and flow rate.

B

When the delivered water properly cleanses the

fixture, replace the Cover Nut and tighten with a

wrench.

DOLPHIN FLUSHOMETER

VALVE ADJUSTMENT

RANGES

FACTORY
SETTING †

MINIMUM
FLUSH
VOLUME

MAXIMUM
FLUSH
VOLUME

CLOSET MODELS: 110, 115, 120,
Type I Class A Shock Proof, and
Type I Class B Rigid Installation

3.5 gpf
(13.2 Lpf)

0.5 gpf
(1.9 Lpf)

16.0 gpf
(60.6 Lpf)

CLOSET MODELS:
111, 115-1.6 and 120-1.6

1.6 gpf
(6.0 Lpf)

0.5 gpf
(1.9 Lpf)

16.0 gpf
(60.6 Lpf)

CLOSET MODEL: 111-1.28

1.28 gpf
(4.8 Lpf)

0.5 gpf
(1.9 Lpf)

16.0 gpf
(60.6 Lpf)

URINAL MODELS: 186-1.0,
Type II Class A Shock Proof, and
Type II Class B Rigid Installation

1.0 gpf
(3.8 Lpf)

0.25 gpf
(0.9 Lpf)

12.0 gpf
(45.4 Lpf)

PUSH AND HOLD

HANDLE DOWN

HANDLE

IN UP

POSITION

TURN CLOCKWISE TO CLOSE

OR COUNTERCLOCKWISE TO

OPEN CONTROL STOP

COUNTERCLOCKWISE

decreases both length of

flush and flush volume

CLOCKWISE increases

both length of flush and

flush volume

† Factory settings are made at 40 psi (275 kPa). Flush volume

will vary with pressure and flow rate.

5 - FLUSH OUT SUPPLY LINE, ADJUST CONTROL STOP, AND INSTALL STOP CAP

6 - ADJUST DURATION OF FLUSH

TROUBLESHOOTING GUIDE

CARE AND CLEANING

DO NOT USE abrasive or chemical cleaners (including chlorine bleach) to clean Flushometers that may dull the luster and attack the chrome or special

decorative finishes. Use

ONLY mild soap and water, then wipe dry with clean cloth or towel.

When cleaning, protect the Flushometer from any splattering of cleaner. Acids and cleaning fluids will discolor or remove chrome plating.

This manual is related to the following products: