Microscope Application Program http://www.ViTiny.com
Copyright ©2008-2012 MicroLinks Technology Corp.
Figure5-41 horizontal scale range
5.60mm
Figure5-42 vertical scale range
4.20mm
EX5: Calibrate the scale for Contacted in high mag.
Use the “calibrator” as a measurement base, flatly put the machine on the
observing object, adjust the focus to find the clearest image. See Fig.5-43 and
Fig.5-44 the result from measurement. Horizontal and vertical scale is 1.40mm
and 1.05mm. Using a calibrator is more easily and accurately to get scale range.
Figure5-43 horizontal scale range
1.40mm
Figure5-44 vertical scale range
1.05mm
(4) Factors cause the inaccurate on the measurement
There are few factors could cause the inaccurate on the measured figures.
The used ruler is not exactly accurate, for example, a metal ruler is not accurate
than a calibrator.
Because of the discrepancy, it is unable to know the accurate horizontal and
vertical scale range, too.
The resolution is not the same with the Windows size.