Specifications, Analog input, Chapter 6 – Measurement Computing WEB-TEMP User Manual
Page 29

29
Chapter 6
Specifications
All specifications are subject to change without notice.
Typical for 25 °C unless otherwise specified.
Specifications in italic text are guaranteed by design.
Analog input
Table 1. Generic analog input specifications
Parameter
Condition
Specification
A/D converters
Four dual 24-bit, Sigma-Delta type
Number of channels
8 differential
Input isolation
500 VDC minimum between field wiring and
power supply input.
Channel configuration
Software programmable to match sensor type
Differential input voltage range
Thermocouple
±0.080 V
RTD
0 V to 0.5 V
Thermistor
0 V to 2 V
Semiconductor sensor
0 V to 2.5 V
Absolute maximum input
voltage
±C0x through ±C7x relative to
AGND
±25 V (power on)
±40 V (power off)
Input impedance
5 GΩ (power on)
1 MΩ (power off)
Input leakage current
Open thermocouple detect disabled 30 nA max
Open thermocouple detect enabled
105 nA max
Common mode rejection ratio
f
IN
= 60 Hz
100 dB min
ADC resolution
24 bits
ADC No missing codes
24 bits
Input coupling
DC
Warm-up time
30 minutes min
Open thermocouple detect
Automatically enabled when the channel pair is
configured for thermocouple sensor.
The maximum open detection time is 3 seconds.
CJC sensor accuracy
15 °C to 35 °C
–0.75 °C to 0.5 °C max
0 °C to 55 °C
–1.5 °C to 1.25 °C max