Hoefer SE300 miniVE User Manual
Page 15

•
p7
Turn each screw (alternating to keep the pressure
even) until the clamps are loosely secured and will
allow the spacers to be adjusted, if necessary. Repeat
on the other side.
If the spacers and glass plates are not perfectly
aligned against the stops, use the stiff end of the
Hoefer Wonder Wedge to press against the edges of
the spacer and glass plates and position them flush
against the guide foot.
Complete clamping by tightening each screw firmly,
hand tight. Do not overtighten, as the plates may
crack. Check the spacer alignment.
Important! Check the alignment of
the bottom edge of the sandwich
against the guide feet (Fig 7).
The spacer must
not protrude out of
the sandwich (or be
recessed into it).
The glass plate should not
be resting on the head of the
spacer “T”.
Fig 8. Misalignments cause leaks.
guide foot
Fig 7. Check the bottom edges for
proper alignment.
- IEF100 (65 pages)
- IEF100 (65 pages)
- SE900 (37 pages)
- SE260 (31 pages)
- SE260 (31 pages)
- SE300 miniVE (37 pages)
- SE300 miniVE (37 pages)
- SE410 (51 pages)
- SE410 (50 pages)
- SE640 (51 pages)
- SE640 (50 pages)
- SE275 (21 pages)
- SE245 (13 pages)
- SE675 (27 pages)
- SG100 (9 pages)
- SG500 (12 pages)
- SQ33 Sequencer (30 pages)
- SQ33 Sequencer (30 pages)
- SQ33 Sequencer (30 pages)
- HE33 (26 pages)
- HE33 (26 pages)
- HE99X (26 pages)
- HE99X (26 pages)
- HE-PLUS System (26 pages)
- HE-PLUS System (26 pages)
- SUB Series (31 pages)
- SUB Series (31 pages)
- TE70X (31 pages)
- TE70X (31 pages)
- PR648 (11 pages)
- PR150 (9 pages)
- PR645 Immunoblot XL (21 pages)
- TE22 (35 pages)
- TE22 (35 pages)
- TE62 (35 pages)
- TE22 (31 pages)
- DQ300 (30 pages)
- DQ300 (30 pages)
- UVIS20 (21 pages)
- UVIS20 (21 pages)
- SP-2001 Vision Life Spectrophotometer (66 pages)
- Single Channel Variable Pipette (8 pages)
- PR250 (18 pages)
- PP24 (10 pages)