beautypg.com

Specifications – Hoefer SE300 miniVE User Manual

Page 11

background image

p3

Specifications

Electrophoresis

Gel sandwich size

10.0 cm wide

×

8 to 10.5 cm long

Max. tank volume

1.6 liters with one module in place
1.4 liters with two modules in place

Max. voltage

600 V~

Max. wattage

25 W per electrophoresis module

Electrotransfer

Max. volume
(blot module)

350 ml per module

Max. tank volume

1.7 liters with one module in place

(for passive cooling)

1.2 liters with two modules in place

Max. wattage

15 W per blot module

Max. current

400 mA

SE300 miniVE specifications

Max. operating temp.

75 °C

Chemical compatibility

For use only with dilute aqueous
solutions between pH 2 and
pH 12. Not compatible with organic
solvents or concentrated alcohols,
acids, bases, and oxidizing agents.

Environmental operating Indoor use: 4 – 40 °C
conditions

Humidity: up to 80%

Altitude: up to 2000 m

Installation category: II

Pollution degree: 2

Dimensions (w

×

d

×

h) 19.2

×

17.2

×

18.8 cm

(7.6

×

6.8

×

7.4 in.)

Weight (tank, lid,

1.2 kg (2.65 lbs)

and two gel modules)
*Product certifications

EN61010–1, UL61010A-1,

CSA C22.2 1010.1, CE Certified

* This declaration of conformity and

the warranty are only valid when the
instrument is used: in laboratory loca-
tions, within the conditions specified
in the user manual, as delivered from
Hoefer, Inc. except for alterations
described in the user manual, and
connected to other CE labeled instru-
ments or products recommended or
approved by Hoefer, Inc.