Specifications – Hoefer SE300 miniVE User Manual
Page 11

•
p3
Specifications
Electrophoresis
Gel sandwich size
10.0 cm wide
×
8 to 10.5 cm long
Max. tank volume
1.6 liters with one module in place
1.4 liters with two modules in place
Max. voltage
600 V~
Max. wattage
25 W per electrophoresis module
Electrotransfer
Max. volume
(blot module)
350 ml per module
Max. tank volume
1.7 liters with one module in place
(for passive cooling)
1.2 liters with two modules in place
Max. wattage
15 W per blot module
Max. current
400 mA
SE300 miniVE specifications
Max. operating temp.
75 °C
Chemical compatibility
For use only with dilute aqueous
solutions between pH 2 and
pH 12. Not compatible with organic
solvents or concentrated alcohols,
acids, bases, and oxidizing agents.
Environmental operating Indoor use: 4 – 40 °C
conditions
Humidity: up to 80%
Altitude: up to 2000 m
Installation category: II
Pollution degree: 2
Dimensions (w
×
d
×
h) 19.2
×
17.2
×
18.8 cm
(7.6
×
6.8
×
7.4 in.)
Weight (tank, lid,
1.2 kg (2.65 lbs)
and two gel modules)
*Product certifications
EN61010–1, UL61010A-1,
CSA C22.2 1010.1, CE Certified
* This declaration of conformity and
the warranty are only valid when the
instrument is used: in laboratory loca-
tions, within the conditions specified
in the user manual, as delivered from
Hoefer, Inc. except for alterations
described in the user manual, and
connected to other CE labeled instru-
ments or products recommended or
approved by Hoefer, Inc.