beautypg.com

The electrophoresis module – Hoefer SE300 miniVE User Manual

Page 12

background image

p4

The Electrophoresis module

This section describes the use of the electropho-
resis module. For instructions on using the blot
module, see page 18.

The electrophoresis module accepts both self-
cast and precast gels 8 cm wide, from 8–10.5 cm
long, and 0.75–1.5 mm thick. For instructions on
using the module with precast gels, see page 10.

Preparing the module

To position the module to accept the gel sand-
wich, each of the three hinged sealing elements
must be opened.

Release the sealing plate by applying gentle inward
pressure to both tabs as indicated by the arrows (Fig 2).

Holding the tabs, move the plate into the fully
open position.

Loosen all four screws 4–5 turns in the counter-
clockwise direction. Do not attempt to remove the
screws from the clamps.

Note: The sealing plate has three
positions: closed, or sealed for
casting; half open, for electropho-
resis; and fully open, for placing
the gel sandwich.

clamp screws (4)

apply inward
pressure to release
sealing plate

Fig 2. Module in the
closed position.